Transverse and slightly oblique extraarticular metacarpal fractures may be fixed with compression plating using a T-plate.
If standard plates are used, bone graft from the distal radius may be necessary to avoid delay in bone healing and consequent secondary displacement, or failure of fixation. Angularly stable plate fixation, using a locking compression plate (LCP), prevents secondary displacement without the need of bone grafting.
Select a plate (1.5–2.0 mm) according to the size of the bone, fracture geometry, and surgeon’s preference. The variable-angle (VA) plate shown on the left can be used for reconstruction of rotational malalignment.
The T-plate is available as a plate with VA locking-head screws. This plate type has the advantage of allowing the insertion of two or three screws at variable angles into the articular block. The rounded plate edges avoid soft-tissue irritation and adhesion.
The plate needs to be contoured to fit the anatomy of the end segment.
This procedure is usually performed with the patient in a supine position with the arm on a radiolucent side table.
For this procedure, the following approaches may be used:
Any axial displacement should be reduced to gain full length of the metacarpal.
Angularly stable plate fixation in the proximal fragment allows for reduction using the plate as a lever (as described below).
Alternatively, and especially when a conventional plate is used, reduction can be performed with axial traction on the thumb, either manually, or with help of a finger trap, and simultaneous pressure on the dorsal aspect of the diaphysis.
Preliminary stabilization can be achieved with an obliquely inserted K-wire.
Be aware that the cortical bone in the metacarpals is dense and thick, so the K-wire tip gets very hot. Irrigation during K-wire insertion is essential to avoid thermal necrosis of the bone around the wire track.
Confirm correct rotational reduction by inspecting the thumb nail plate in relation to the index finger. Compare it to the contralateral side if necessary.
Adapt the plate length to the length of the metacarpal. Avoid sharp edges, which may be injurious to the tendons.
There should be at least 3 plate holes distal to the fracture available for fixation in the diaphysis. At least two screws need to be inserted into the diaphysis.
The plate must be adapted accurately to the contour of the proximal fragment. Use bending pliers to contour the plate. Be careful to bend in between the holes by positioning the pliers over the plate holes. Alternatively, insert the threaded bolt into the plate holes to bend the plate.
The bend should never go through a hole as this may damage the locking mechanism.
To prevent the screw from penetrating the joint, a slight bend may be introduced to the articular part of the plate. Be careful not to overbend, as there is a comminuted zone at the opposite cortex, which cannot withstand compression.
Alternatively, if the fracture allows, the plate may be positioned more distally.
Start with inserting screws into the articular block:
Mark the plane of the 1st CMC joint with a thin K-wire or a hypodermic needle. Put the plate on the dorsal aspect of the proximal fragment, with a threaded drill guide in one of the outer holes. Drilling must be parallel to the marker wire (or needle) to ensure that the screw will not penetrate the joint. Confirm using image intensification.
Insert the locking screw without completely tightening it.
Align the plate along the axis of the shaft.
Insert a second screw using the same technique. Always use a threaded drill guide when preparing for locking screws.
If possible, add another screw in the transverse part of the T-plate.
Pressing the plate down to the diaphyseal fragment with a periosteal elevator will automatically tilt the articular fragment into its correct position. Check the reduction of the dorsal fracture line.
Confirm with an image intensifier.
Insert a screw through the most distal plate hole. This screw would normally be a standard screw, as it engages in thick cortical bone and pulls the distal part of the plate to the bone.
The second distal screw is inserted using the same technique. This may be either a cortical screw or a locking-head screw as required.
If compression plating can be applied (in case of minimal comminution), insert a cortical screw in compression mode in the oblong hole.
Confirm correct rotational alignment by clinical examination.
Image intensification may be used to confirm anatomical reduction and correct placement of implants in two views.
The aftercare can be divided into four phases of healing:
Full details on each healing phase can be found in the chapter Postoperative treatment of thumb injuries.
Early range of motion, without load bearing, in the hand is important for optimal outcome.
Immediately postoperatively, a temporary plaster splint is applied. This also helps with swelling and pain control.
The plaster splint can be switched to a custom-made thermoplastic splint after a week. The splint should be worn until fracture healing has been confirmed.
Hand therapy should be initiated as soon as possible.
The arm should be actively elevated to help reduce any swelling.
To prevent joint stiffness, the patient should be instructed to begin active motion (flexion and extension) of all nonimmobilized joints of the hand immediately after surgery.
In compliant patients with stable fixation, the splint can be removed under the supervision of a hand therapist, and early active motion exercises started after a few days.
Weight bearing should be avoided until fracture healing is confirmed.
The surgical wound is checked as necessary depending on soft tissue status. The wound is monitored for evidence of infection.
To allow for early joint mobilization and to prevent wound dehiscence, leave the sutures for 2 weeks.
Confirm fracture healing clinically and radiologically after about 5–6 weeks (depending on the fracture configuration).
The implants may need to be removed in cases of soft-tissue irritation.
In case of joint stiffness or tendon adhesion restricting finger movement, arthrolysis or tenolysis may become necessary. In these circumstances, the implants can be removed at the same time.