1. Home
  2. Skeleton
  3. Diagnosis
  4. Indications
  5. Treatment

Authors of section

Authors

Aida Garcia, Fabio A Suarez

Executive Editor

Simon Lambert

Open all credits

Lag-screw and plate fixation

1. General considerations

Introduction

A wedge fragment may be fixed to both end segments with lag screws if the fractures are sufficiently oblique and the wedge fragment large enough. A lateral or dorsal neutralization plate is then applied to protect the lag screw.

This technique is useful to gain more stability if particularly only nonlocking plates are available.

If the fracture morphology allows only one lag screw, fracture stabilization is completed with a lateral or dorsal bridge plate.

Smaller lag screws (1.0 or 1.3 cortical) should be used to avoid vascular compromise and further fragmentation. In doubt, bridge plating should be considered.

Lag-screw and plate fixation of a wedge fracture of the proximal phalangeal shaft

Lag-screw planning

Obliquity of the wedge fracture is possible either in the plane visible in the AP or lateral views. Always confirm the fracture configuration with views in both planes.

In this procedure, an oblique wedge fracture visible in the lateral view is shown with lag-screw fixation protected with a lateral plate.

Wedge fracture of the proximal phalangeal shaft

Plate selection

For plating, several plate types may be used:

  • Conventional straight locking compression plate (LCP); dorsal and lateral
  • T-plate; dorsal
  • Strut plate; dorsal
  • Condylar plate; lateral or dorsal
  • Phalangeal base plate; lateral or dorsal

The plates may come with or without variable-angle (VA) locking-head screws. If an anatomical plate is not available, a conventional minicondylar plate may be used.

The plate selection depends on the fracture pattern and should allow at least two screws in the proximal and distal main fragment.

In this procedure, plate fixation with a lateral phalangeal base plate is shown.

Selection of plate types for fixation of proximal phalangeal fractures

2. Patient preparation

Place the patient supine with the arm on a radiolucent hand table.

Patient positioned supine with the arm on a radiolucent hand table

3. Approaches

For this procedure, the following approach is normally used for application of a lateral plate:

Application of a dorsal plate may require a dorsal approach:

4. Reduction

Gaining length

Length can be gained by traction applied manually, by a finger trap, or with pointed reduction forceps.

Confirm length and rotation with the neighboring fingers with the metacarpophalangeal (MCP) joint in 90° flexion.

Reduction of a wedge fracture of the proximal phalangeal shaft

Reducing rotation and angulation

Manipulate the fragments with pointed reduction forceps to reduce rotation and restore correct angulation.

Pitfall: excessive pressure may cause fracture comminution.
Holding a reduction of a wedge fracture of the proximal phalangeal shaft with reduction forceps

Preliminary fixation

Preliminarily secure the reduction with K-wires or reduction forceps.

Holding a reduction of a wedge fracture of the proximal phalangeal shaft with rK-wires or eduction forceps

Alternatively, hold the reduction with reduction forceps designed for percutaneous technique. Impingement of soft tissues should be avoided.

Holding a reduction of a wedge fracture of the proximal phalangeal shaft with percutaneous reduction forceps

5. Checking alignment

Identifying malrotation

At this stage, it is advisable to check the alignment and rotational correction by moving the finger through a range of motion.

Rotational alignment can only be judged with flexed metacarpophalangeal (MCP) joints. The fingertips should all point to the scaphoid.

Malrotation may manifest by an overlap of the flexed finger over its neighbor. Subtle rotational malalignments can often be judged by a tilt of the leading edge of the fingernail when the fingers are viewed end-on.

If the patient is conscious and the regional anesthesia still allows active movement, the patient can be asked to extend and flex the finger.

Any malrotation is corrected by direct manipulation and later fixed.

73 P130 Lag screw fixation

Using the tenodesis effect when under anesthesia

Under general anesthesia, the tenodesis effect is used, with the surgeon fully flexing the wrist to produce extension of the fingers and fully extending the wrist to cause flexion of the fingers.

Surgeon fully flexing the wrist to produce extension of the fingers and fully extending the wrist to cause flexion of the fingers

Alternatively, the surgeon can exert pressure against the muscle bellies of the proximal forearm to cause passive flexion of the fingers.

Surgeon exerting pressure against the muscle bellies of the proximal forearm to cause passive flexion of the fingers

Confirm fracture fixation and stability with an image intensifier.

6. Lag-screw fixation

The lag screw should be inserted centered on and as perpendicularly to the fracture plane as possible and from the dorsal surface. The direction of the obliquity of the fracture plane dictates the exact position of the lag screw.

Note: Countersinking in the diaphysis should be performed with care as it risks iatrogenic fractures.
Lag-screw fixation of a wedge fracture of the proximal phalangeal shaft

Fix the wedge fragment to the other end segment with another lag screw if the size of the wedge fragment and the fracture plane allow it.

Lag-screw fixation of a wedge fracture of the proximal phalangeal shaft

Screw size selection

The exact size of the diameter of the screws used will be determined by the fragment size and the fracture configuration.

The various gliding and thread hole drill sizes for different screws are illustrated here.

Screw sizes of the diameters and thread hole drill sizes

Screw length pitfalls

Ensure that a screw of the correct length is used.
  • Too short screws do not have enough threads to engage the cortex properly. This problem increases when self-tapping screws are used due to the geometry of their tip.
  • Too long screws endanger the soft tissues, especially tendons and neurovascular structures. With self-tapping screws, the cutting flutes are especially dangerous, and great care has to be taken that the flutes do not protrude beyond the cortical surface.
Correct screw length for fracture fixation means screws should not be too long or too short.

Pitfall: screw too close to the fracture

Do not insert screws too close to the fracture apex. A minimal distance from the fracture line, equal to the screw head diameter, must be observed.
Screw placement not close to the fracture apex and with minimal distance to the fracture line equal to the diameter of the screw head

Pitfall: beware of fissure lines

Often there are short fissure lines that are not apparent on the x-rays. Check for these under direct vision and ensure the screws are not inserted through these fissure lines.
Screws inserted to avoid passing through short fissures

7. Plate fixation

Neutralization plate

Apply the neutralization plate depending on the fracture plane and the lag-screw position, either dorsally or laterally.

Insert at least two screws proximally and distally to the fracture in neutral mode.

Plate trimming

Select or adapt the plate length to fit the length of the proximal phalanx. Avoid sharp edges, which may be injurious to the tendons.

Plate trimming for fixation of a proximal phalangeal fracture

Contouring

Contour the plate to the shape of the proximal phalanx with the specific bending plyers. This ensures that the plate does not produce displacement of the fracture.

Plate contouring for fixation of a proximal phalangeal fracture

Plate positioning

Place the plate dorsal centered on the diaphysis in the coronal plane, allowing at least two screws in both main fragments.

Lag-screw and plate fixation of a wedge fracture of the proximal phalangeal shaft – plate positioning

Insertion of screws

Insert screws into the proximal main fragments.

Take care not to violate the joint surface.

Confirm reduction and implant position with an image intensifier.

Complete the plate fixation by introducing screws into the other main fragment.

Note: Avoid screw protrusion through the far cortex, as ligament injury may result from friction during movement.
Lag-screw and plate fixation of a wedge fracture of the proximal phalangeal shaft

Bridge plate

If the wedge fragment can only be fixed to one end segment, apply the plate in bridging mode. Compression may be applied with caution and if the fracture line stability allows.

Cover the plate with periosteum to avoid adhesion between the tendon and the implant leading to limited finger movement.

Lag-screw and plate fixation of a wedge fracture of the proximal phalangeal shaft

8. Final assessment

Confirm fracture reduction and stability and implant position with an image intensifier.

Intraoperative image of lag-screw and plate fixation of a wedge fracture of the 5th proximal phalangeal shaft

9. Aftercare

Postoperative phases

The aftercare can be divided into four phases of healing:

  • Inflammatory phase (week 1–3)
  • Early repair phase (week 4–6)
  • Late repair and early tissue remodeling phase (week 7–12)
  • Remodeling and reintegration phase (week 13 onwards)

Full details on each phase can be found here.

Postoperative treatment

If there is swelling, the hand is supported with a dorsal splint for a week. This would allow for finger movement and help with pain and edema control. The arm should be actively elevated to help reduce the swelling.

The hand should be splinted in an intrinsic plus (Edinburgh) position:

  • Neutral wrist position or up to 15° extension
  • MCP joint in 90° flexion
  • PIP joint in extension
Dorsal splint to treat a dislocation of the proximal interphalangeal joint

The reason for splinting the MCP joint in flexion is to maintain its collateral ligament at maximal length, avoiding scar contraction.

PIP joint extension in this position also maintains the length of the volar plate.

73 P130 Lag screw fixation

After subsided swelling, protect the digit with buddy strapping to a neighboring finger to neutralize lateral forces on the finger.

Buddy strapping avoiding direct skin contact with adjacent fingers as conservative treatment

Functional exercises

To prevent joint stiffness, the patient should be instructed to begin active motion (flexion and extension) immediately after surgery.

Functional exercises for the hand

Follow-up

See the patient after 5 and 10 days of surgery.

Implant removal

The implants may need to be removed in cases of soft-tissue irritation.

In case of joint stiffness or tendon adhesion restricting finger movement, arthrolysis or tenolysis may become necessary. In these circumstances, the implants can be removed at the same time.

10. Case

AP x-ray of a wedge fracture of the 5th proximal phalangeal shaft

73 D320 Definition

Lateral x-ray

x-ray of a wedge fracture of the 5th proximal phalangeal shaft

Intraoperative image of lag-screw fixation

Intraoperative image of lag-screw fixation of a wedge fracture of the 5th proximal phalangeal shaft

Intraoperative image of the same case with the plate applied and fixed with the first screws

Intraoperative image of lag-screw and plate fixation of a wedge fracture of the 5th proximal phalangeal shaft

Final construct

Intraoperative image of lag-screw and plate fixation of a wedge fracture of the 5th proximal phalangeal shaft