Revision of a standard stem

1. Principles

The primary principle is to obtain a stable humeral component while creating the environment for the periprosthetic fracture to unite.

ORIF of the humerus is required in addition to revision of the humeral component. Further information can be found in the AO Surgery Reference adult trauma Humeral shaft module.

Preoperative planning ensures:

  • Adequate surgical exposure
  • The necessary inventory of equipment and implants

This illustration shows the final construct after revision of a standard implant to a long implant.

per10 P420 Revision of a short stem

2. Preparation

The standard patient position is the beach chair position with inclination of about 30°. An arm holder may be helpful but is not essential.

Intraoperative fluoroscopy can be helpful.

Patient positioning should be discussed with the anesthetist.

per 10 Pr110 Beach chair position at thirty degrees

3. Approach

The standard approach is a deltopectoral approach.

This may be extended distally or proximally as needed.

per10 P450 Humeral revision with allograft

4. Implant removal

Standard stem prosthesis

In the case of a standard stem prosthesis, osteotomes are used to liberate the periphery of the prosthesis. Implant-specific tools are used to remove the prosthesis.

A humeral osteotomy may be needed to remove the humeral component.

Periprosthetic shoulder fracture - revision of a standard stem implant – implant removal
Pearl: Consider provisional reduction and holding the fracture to facilitate removal of the implant without further fracture displacement. In this case a C-clamp has been applied. An alternative is to apply a cerclage compression wire around the fracture.
Periprosthetic shoulder fracture - revision of a standard stem implant – provisional reduction and holding of the fracture

5. Decision making

Glenoid condition, anatomic configuration

Revision of a standard-stemmed anatomic shoulder replacement to another anatomic configuration is reasonable if:

  • The existing glenoid component is intact and well-fixed
  • The rotator cuff is intact and functioning

If the rotator cuff is not intact, or cannot be repaired, then revision to a reverse total shoulder replacement is preferred.

View of shoulder showing standard stem prosthesis and subscapularis tendon

Glenoid condition, reverse configuration

Revision of a standard-stemmed reverse shoulder replacement to a long-stemmed reverse shoulder replacement is reasonable if:

  • The existing glenoid components are intact and well-fixed

If the existing glenoid components are not secure, then revision in a one- or two-stage procedure with glenoid reconstruction or custom prosthesis may be preferred.

6. Implant removal

Standard stem prosthesis

An implant-specific tool is used to remove the head component.

An implant-specific tool is used to remove the head component of a standard stem prosthesis

Flexible osteotomes may be helpful to disrupt the implant-bone or cement-bone interface.

Flexible osteotomes may be helpful when removing a standard stem prosthesis

The implant may be removed using specific instruments. Alternatively, the implant may be disimpacted by striking the component in an axially aligned manner.

A standard stem prosthesis may be disimpacted by striking the component in an axially aligned manner

Additional humeral osteotomy

Sometimes, humeral osteotomy is required to disrupt the implant-bone or cement-bone interface to facilitate removal of the implant.

Sometimes, humeral osteotomy is required to remove a standard-stem component

The humeral component is removed.

The standard stem component is removed
Pearl: Consider provisional reduction and holding the fracture to facilitate removal of the implant without further fracture displacement. In this case a C-clamp has been applied. An alternative is to apply a cerclage compression wire around the fracture.
Provisional reduction and holding the fracture may facilitate removal of the implant without further fracture displacement

7. Reduction

The fracture should be appropriately exposed.

Bone fragments are reduced to realign the metaphysis and diaphysis.

Periprosthetic shoulder fracture – exposing the fracture

Provisional reduction is achieved with reduction clamps.

Periprosthetic shoulder fracture - provisional reduction is achieved with reduction clamps

In simple fracture patterns, cerclage cables or wires can be used to maintain the reduction.

Pitfall: Extreme care must be taken on passing these wires to avoid damaging the axillary and radial nerves.
per10 P440 Revision of a standard stem

In complex fracture patterns, bridging plate fixation is required to maintain reduction and stability.

Provisional reduction of the fracture may be provided by cerclage wire or sutures.

The appropriate plate for bridging fixation may be applied and provisionally fixed proximally and distally, taking care not to obstruct the implantation of a planned humeral component. Provisional monocortical screws are recommended.

Periprosthetic shoulder fracture – in complex fracture patterns, bridging plate fixation is required to maintain reduction and stability

8. New stemmed implant – preparation and insertion

After reduction and fixation of the fracture the choice of cemented or uncemented revision stemmed implant is based on a number of factors including:

  • Predicted implant-bone interface quality
  • Humeral shaft morphology
  • Bone quality

The implantation technique options include:

  • Uncemented
  • Cemented
  • Hybrid

Uncemented

The uncemented implantation technique is chosen when predicted implant-bone interface quality is sufficient for primary humeral implant stability.

A long stem humeral component is implanted following appropriate preparation, according to the manufacturer's instructions.

Information about the uncemented basic technique is provided here.

After implantation of the humeral component definitive bridging fixation of the fracture is completed using a combination of techniques including short locking screws proximally, supplementary cerclage wires or sutures, and long locking screws distally.

Information about plate fixation principles can be found here.

per10 P420 Revision of a short stem

Cemented

The cemented implantation technique is chosen when predicted implant-bone interface quality is not sufficient for uncemented primary humeral implant stability.

Information about the cemented basic technique is provided here.

After implantation of the humeral component definitive bridging fixation of the fracture is completed using a combination of techniques including short locking screws proximally, supplementary cerclage wires or sutures, and long locking screws distally.

Information about plate fixation principles can be found here.

Pearl: A distal intramedullary cement restrictor should be used to avoid difficulty in achieving distal screw fixation caused by excess distal cement.
Pitfalls:
  • Cement extrusion at the fracture site may compromise eventual fracture healing
  • Cement extrusion at the fracture site may damage the radial nerve (and other close nerves) if not noted
Cemented implantation technique

Hybrid

The hybrid implantation technique is chosen when predicted implant-bone interface quality at, and proximal to, the fracture site is considered likely to heal well enough for secondary humeral implant stability. Cementation of the humeral implant distal to the fracture provides initial humeral stem stability, which then facilitates secondary humeral implant stability as the fracture heals.

Information about the hybrid basic technique is provided here.

After implantation of the humeral component definitive bridging fixation of the fracture is completed using a combination of techniques including short locking screws proximally, supplementary cerclage wires or sutures, and long locking screws distally.

Information about plate fixation principles can be found here.

Pearl: A distal intramedullary cement restrictor should be used to avoid difficulty in achieving distal screw fixation caused by excess distal cement.
Pitfalls:
  • Cement extrusion at the fracture site may compromise eventual fracture healing
  • Cement extrusion at the fracture site may damage the radial nerve (and other close nerves) if not noted
per10 P420 Revision of a short stem

9. Shoulder joint stability

In the setting of an anatomic total shoulder replacement, the rotator cuff must be robust and intact, and the subscapularis tendon must be repaired. If these conditions are not met, a reverse total shoulder replacement should be considered.

Stability of a reverse total shoulder replacement can be improved by:

  • Repair of the greater and/or lesser tuberosities
  • Repair of the subscapularis tendon
  • Using a retentive polyethylene humeral liner
  • Revising the glenosphere
per10 P420 Revision of a short stem

10. Stability verification

Stability of the total shoulder replacement and fracture fixation is assessed by passive movement of the limb through a functional range of movement to include internal rotation (placing the hand by the ipsilateral hip), elevation (placing the hand on the vertex of the head), and external rotation in mid-range elevation (reaching forward).

If necessary, image intensification whilst performing these movements may be helpful.

11. Aftercare

Postoperative phases

The aftercare can be divided into four phases of healing:

  • Inflammatory phase (week 1–3)
  • Early repair phase (week 4–6)
  • Late repair and early tissue remodeling phase (week 7–12)
  • Remodeling and reintegration phase (week 13 onwards)

Full details on each phase can be found here.