Open reduction, K-wire fixation

1. Preliminary remarks

Introduction

Children's hip fractures have historically had a bad prognosis. Accurate stable fixation has been one of the factors identified to improve the outcome.

Stable fractures have a better prognosis than displaced fractures.

The main problems have been avascular necrosis (AVN) due to disruption of the blood supply, nonunion and malunion due to unstable fixation.

The objective is timely fixation without disruption of the blood supply.

open reduction k wire fixation

Equipment selection

2.0, 2.5, or 3.0 mm K-wires can be selected depending upon the size of the child.

Larger K-wire sizes are preferred for stability.

Threaded K-wires also increase stability.

open reduction k wire fixation

Implant insertion

Typically, two K-wires separated vertically are inserted. This configuration may be more suitable in smaller bones.

An alternative is three K-wires used in an apex-distal, triangular configuration, with the lower central K-wire abutting the calcar.

Inserting a single central K-wire does not provide rotational stability.

open reduction k wire fixation

K-wires alone may not provide adequate stability for a vertically-oriented fracture line. A pediatric hip locking plate is a better construct if the fracture line is more vertical.

If the child is too small for a plate, the K-wires must be protected with a hip spica cast.

open reduction k wire fixation

2. Patient preparation and approaches

Patient preparation

This procedure is normally performed with the patient in a supine position.

in situ fixation with k wires or screws

Approaches

For this procedure the following approaches may be used:

3. Reduction

A K-wire is inserted through the lateral cortex of the femur just above the level of the lesser trochanter and advanced to the fracture line.

If necessary, manipulation of the femoral head is achieved using a small K-wire, inserted into the proximal fragment, as a joystick.

Pitfall: Secondary subtrochanteric fractures may occur if K-wires are inserted below the level of the lesser trochanter.

open reduction k wire fixation

The femoral neck is manipulated gently into position, usually by lifting with a bone hook and manipulating the K-wires.

Gentle reduction maneuvers are essential in order to prevent secondary damage to the femoral head blood supply.

open reduction k wire fixation

4. Fixation

K-wire insertion

When acceptable reduction has been obtained, the first K-wire is advanced across the fracture and into the femoral head. Overpenetration into the hip joint must be avoided.

open reduction k wire fixation

Additional wires are inserted under x-ray control, using image intensification.

open reduction k wire fixation

The K-wire positions are confirmed on AP and lateral x-ray images, using image intensification.

open reduction k wire fixation

If using a standard radiolucent table, the K-wire fixation is generally stable enough to allow a “frog” lateral view.

The wires are repositioned if necessary.

in situ fixation with k wires or screws

Images at multiple angles are used to confirm that the K-wires do not penetrate into the hip joint.

in situ fixation with k wires or screws

Pitfall: Wire penetration can occur, especially with eccentrically placed wires, even if not apparent on standard AP or lateral x-rays.

Typically, 5 mm of epiphyseal bone/cartilage should remain between the wire tip and joint surface.

Dynamic, real-time image intensification, with a full range of internal and external rotation at different degrees of flexion, is useful to see how close the K-wire-tip is to the joint surface of the head.

Observing an approach/withdrawal of the K-wire tip helps in judging its position. Some surgeons supplement this examination with 3-D image intensification for confirmation.

Arthrography is useful to confirm correct wire placement in the younger patient.

open reduction k wire fixation

Cutting the K-wires

The K-wires can be bent and cut short and left beneath the fascia lata.

The K-wires are usually removed with a short secondary procedure once the fracture is healed.

31 m 31 i

5. Hip spica

Because K-wire fixation is less stable, and only indicated for younger children, a hip spica cast is generally used to supplement the fixation.

Click here for details of hip spica application.

31 e 11

6. Aftercare

Introduction

Hip spica is only likely to be used in the treatment of small children with proximal femoral fractures.

For larger children, internal fixation should be used even for nondisplaced fractures.

31 e 11

Immediate care

After application, the spica should be trimmed to allow adequate space for bodily functions. The edges of the spica should be padded and waterproofed.


Diaper care

Generally, a hip spica should allow space for a small diaper inside the plaster and a large one outside the plaster. The diapers should be positioned to prevent soiling of the spica.

Washing

The spica is not waterproof. Bathing and showering should not be attempted. Hair washing should be done very carefully.

Skin care

No skin products should be put inside the spica. Only skin that can be seen should be moisturized.

Transport

Both the child and the spica must be lifted. A special car seat will be required. The child may be placed in a stroller or buggy.

In this circumstance, the parents/carers are advised to return to the healthcare provider.

Length of time in spica

The length of time in the spica depends on the age of the child and the healing of the fracture. A proximal femoral fracture in a child aged below 4 years should always be healed in 6 weeks.

Growth

The child will continue to grow and the tightness of the spica should be monitored.

Follow-up x-rays

Nondisplaced fractures being treated nonoperatively should have early radiological follow-up. If the spica is being used for protection of fixation, x-rays are required only when planning removal.