1. Home
  2. Skeleton
  3. Diagnosis
  4. Indications
  5. Treatment

Plate fixation of a complete articular fracture

1. General considerations

Introduction

Articular fractures, simple or comminuted, of a metacarpal head may be fixed with a T-, Y-, or an anatomical neck plate.

The articular fragments are reduced and fixed using the transverse part of the plate. In a second step, the extraarticular fracture component is reduced and fixed.

Noncomminuted fracture components, eg, between the articular fragments or between the articular block and the diaphysis, may be fixed with compression.

The reduction can be assisted with arthroscopy if skill and equipment are available.

Complete articular fracture of the distal end segment of the 1st metacarpal - Plating

Plate selection

Select a plate (1.5–2.0 mm) according to the size of the bone, fracture geometry, and surgeon’s preference. A T-shaped plate may be used. Other shapes may also be applied.

A plate with variable-angle (VA) locking-head screws has the advantage to allow insertion of two or three screws in variable angles into the articular block. The rounded plate edges avoid soft-tissue irritation and adhesion.

T- and Y-shaped VA locking plates and metacarpal neck plates

2. Patient preparation

This procedure is usually performed with the patient in a supine position with the arm on a radiolucent side table.

Patient in supine position with the arm on a side table

3. Approach

For this procedure, the following approaches may be used:

Dorsal approach to the metacarpophalangeal joint of the thumb

4. Reduction

Manipulate the articular fragments with a dental pick, small K-wires, or a small periosteal elevator.

Small K-wires can also be used for preliminary fixation. Depending on fracture configuration, pointed forceps may be useful for reduction.

Complete articular fracture of the distal end segment of the 1st metacarpal – Reduction and preliminary K-wire stabilization

Management of impaction

In impacted fractures, reduce the articular surface, and fill the bony defect under the fragments with bone graft. This helps to keep the fragments in place when the screw fixation is performed.

Complete articular fracture of the distal end segment of the 1st metacarpal – Management of impaction

Buttressing small fragments

Additionally, a periarticular K-wire may be used to reposition and support impacted articular fragments.

The end of the K-wire may protrude through the skin, which facilitates removal after consolidation of the fracture.

Complete articular fracture of the distal end segment of the 1st metacarpal – K-wire stabilization of the articular block before plate application

Confirming anatomical reduction

Check anatomical reduction of the joint surface under direct view or arthroscopy and image intensification.

Maximally flex the MCP joint to gain a view of the palmar aspect of the metacarpal head.

5. Checking alignment

Confirm correct rotational reduction by inspecting the thumb nail plate in relation to the index finger. Compare it to the contralateral side if necessary.

Checking correct rotational alignment of a fracture in the thumb by comparing the nail orientation with the uninjured thumb

6. Screw fixation of the articular block

If the fracture line allows, eg, in a T- or Y-shaped fracture with the articular fracture line running from palmar to dorsal, a lag screw may be inserted to stabilize the articular fracture. This converts a three-part fracture to a two-part fracture.

Make sure the screw track is not interfering with later plate application.

Complete articular fracture of the distal end segment of the 1st metacarpal – Percutaneous lag-screw fixation of the articular block before plate application

7. Plate preparation

Plate trimming

Adapt the plate length to the length of the metacarpal. Avoid sharp edges, which may be injurious to the tendons. There should be at least 3 plate holes proximal to the fracture available for fixation in the diaphysis. At least two screws need to be inserted into the diaphysis.

Plate trimming with special tool

Contouring

The plate must be adapted accurately to the contour of the proximal fragment. Use bending pliers to contour the plate. Be careful to bend in between the holes by positioning the pliers over the plate holes. Alternatively, insert the threaded bolt into the plate holes to bend the plate.

The bend should never go through a hole as this may damage the locking mechanism.

Contouring the T-shaped end of the plate to the bone

8. Bridge plating

Plate application

Place the plate exactly so that the articular fragments can be fixed using screws through the distal plate holes. Often, it is not possible to insert a screw into each fragment. Small fragments should be supported by adjacent large fragments or bone graft.

Hold the plate preliminarily with a cortical screw, not fully tightened, in the oblong hole. This allows for later adjustment of the plate position.

Complete articular fracture of the distal end segment of the 1st metacarpal - Application of a T-plate

If using a metacarpal neck plate, position it with the distal end proximal to the cartilage, if possible proximal to the joint capsule.

Complete articular fracture of the distal end segment of the 1st metacarpal – Application of a metacarpal neck plate

If necessary, adjust the plate position.

Begin fixation of the articular block with the most critical articular fragment.

Drill carefully not to displace the fragments and penetrate the articular surface.

Insert the first screw.

Confirm the reduction and correct screw position using image intensification.

Complete articular fracture of the distal end segment of the 1st metacarpal – aligning the plate with the bone axis

Insertion of further diaphyseal screws

Insert the second and, if possible, a third screw into the diaphyseal fragment and tighten them.

When the fixation is completed, remove any K-wires, other than those needed to buttress small articular fragments.

Cover the plate with periosteum, if possible, to avoid adhesion between the tendon and the implant leading to limited finger movement.

Complete articular fracture of the distal end segment of the 1st metacarpal - T-plate fixation

9. Compression plating

Compression of articular fracture

If the articular fracture allows, apply extrinsic compression.

Compress the fracture with reduction forceps. Position one tip of the forceps on the plate with the first screw to avoid damage of the bone cortex.

Insert the second VA locking screw to hold the compression.

Complete articular fracture of the distal end segment of the 1st metacarpal - Extrinsic compression of the simple articular fracture during T-plate application

Compression of metadiaphyseal fracture

If the metadiaphyseal fracture plane allows, the plate may be applied in compression mode.

After fixation of the articular block to the plate with two screws, insert a cortical screw in the oblong hole in compression mode to compress the extraarticular fracture component.

Check again rotational alignment clinically and with an image intensifier.

Insert a second screw into the diaphyseal segment to finalize the fixation.

Complete articular fracture of the distal end segment of the 1st metacarpal – Compression plating

10. Final assessment

Confirm correct rotational alignment by clinical examination.

Image intensification may be used to confirm anatomical reduction and correct placement of implants in two views.

11. Aftercare

Healing phases

The aftercare can be divided into four phases of healing:

  • Inflammatory phase (week 1–3)
  • Early repair phase (week 4–6)
  • Late repair and early tissue remodeling phase (week 7–12)
  • Remodeling and reintegration phase (week 13 onwards)

Full details on each healing phase can be found in the chapter Postoperative treatment of thumb injuries.

Postoperative care

Early range of motion, without load bearing, in the hand is important for optimal outcome.

Immediately postoperatively, a temporary plaster splint is applied. This also helps with swelling and pain control.

The plaster splint can be switched to a custom-made thermoplastic splint after a week. The splint should be worn until fracture healing has been confirmed.

Hand therapy should be initiated as soon as possible.

The arm should be actively elevated to help reduce any swelling.

Thumb spica splint immobilizing wrist, CMC, and MCP joints of the thumb and custom-made thermoplastic splint

Mobilization

To prevent joint stiffness, the patient should be instructed to begin active motion (flexion and extension) of all nonimmobilized joints of the hand immediately after surgery.

In compliant patients with stable fixation, the splint can be removed under the supervision of a hand therapist, and early active motion exercises started after a few days.

Weight bearing should be avoided until fracture healing is confirmed.

Functional exercises for mobilization of the hand

Follow-up

The surgical wound is checked as necessary depending on soft tissue status. The wound is monitored for evidence of infection.

To allow for early joint mobilization and to prevent wound dehiscence, leave the sutures for 2 weeks.

Confirm fracture healing clinically and radiologically after about 5–6 weeks (depending on the fracture configuration).

Implant removal

The implants may need to be removed in cases of soft-tissue irritation.

In case of joint stiffness or tendon adhesion restricting finger movement, arthrolysis or tenolysis may become necessary. In these circumstances, the implants can be removed at the same time.