Authors of section

Authors

Andrew Howard, Theddy Slongo, Peter Schmittenbecher

Executive Editor

James Hunter

General Editor

Fergal Monsell

Open all credits

Monteggia equivalent, oblique fracture of ulna: open reduction, plate and screw fixation

1. Introduction

General considerations

Plating is the standard technique for treating forearm fractures in adults and is therefore best considered for skeletally mature or nearly mature children.

Children with open physes have thick active periosteum favoring stability and rapid healing with the ESIN method. Where such techniques are unavailable plating may be used in younger children.

If technically possible, ESIN is biologically favored. If plating is used, soft-tissue and periosteal stripping should be minimized.

Plating of oblique or spiral forearm shaft fractures

Plating of pediatric forearm shaft fractures follows the technique for plate fixation in adults.

For oblique or spiral forearm shaft fractures, interfragmentary compression can be achieved by one of the following techniques:

  • Lag screw outside the plate
  • Lag screw through the plate
  • Compression plate with consecutive lag screw through the plate

This results in absolute fracture stability.

Large wedge fragments may be fixed to one of the main segments before plate fixation with a lag screw, placed through or outside the plate.

For unstable multifragmentary forearm shaft fractures, bridge plating may be applied. This results in relative fracture stability.

Plating of oblique or spiral forearm shaft fractures

2. Principles

Instruments and implants

A small fragment set consists of:

  • 2.7 or 3.5 mm plates and screws
  • Power driver
  • 2.7 or 3.5 mm insertion set
Instruments and implants

Compression by lag screw

For interfragmentary compression a lag screw can be inserted either through a plate hole (1) or separate from the plate (2).

Compression by lag screw

Compression plating

Axial compression results from eccentric screw (load screw) insertion with compression plates.

Compression plating

Plate position

The ulnar plate can be positioned under the extensor carpi ulnaris (A), under the flexor carpi ulnaris (B), or in the interval between extensor and flexor carpi ulnaris (C).

With a more proximal fracture it can be more difficult to place the plate underneath the extensor muscle.

With more distal fractures, plates should be placed in the interval between the extensor and flexor carpi ulnaris (C).

Plating the tensile (subcutaneous) aspect of the ulna (C) is biomechanically preferable, but plate prominence can cause irritation.

Plates in positions A or B are covered by the muscle compartment and avoids this complication.

In the following procedure, the plate positioned on the subcutaneous border of the ulna (C) is illustrated.

Plate position

Approach

The ulna is exposed by the direct approach between the flexor and extensor muscle compartments.

open reduction plate fixation

3. Reduction

Reduction and fracture stability in children

Children’s periosteum is thick, tough tissue and is often intact on the concave (compression) side of a fracture.

Reduction maneuvers should be gentle to take advantage of the stability of the intact periosteum.

Exaggeration of the fracture deformity may be required to loosen the periosteum and allow for gentle reduction.

In pediatric fractures there is often a combination of patterns of bone failure. Residual plastic deformity may prevent anatomical reduction of some of the fracture edges. Provided the alignment of the bone is anatomical and overall reduction is stable it is not necessary to perfectly reduce the entire fracture.

Periosteum as a factor in reduction

Open reduction

Reduce the fracture anatomically, using a reduction forceps on each main fragment.

The use of blunt, as opposed to pointed, reduction forceps can be helpful, particularly if greater forces are required.

Open reduction with forceps

Pearl: twisting a reduction forceps

Reduction of overlapping oblique fractures can be achieved by twisting a reduction forceps, thereby lengthening the fracture.

Open reduction by twisting a reduction forceps

Maintaining fracture reduction

Maintain the fracture reduction with pointed reduction forceps.

Place the forceps such that it will not interfere with the planned plate position.

Open reduction by twisting a reduction forceps

4. Plate length and number of screws

In the forearm, in addition to any lag screw, three bicortical plate screws are suggested in each main fracture fragment due to the high torsional stresses.

Not every plate hole needs to be occupied by a screw.

In oblique or spiral forearm shaft fractures an empty plate hole may be necessary at the level of the fracture to introduce a lag screw.

Open reduction by twisting a reduction forceps

5. Fixation – lag screw as primary fixation device outside plate

Principle

If the fracture morphology dictates that the interfragmentary lag screw be positioned separate from the plate, the plate then acts as a neutralization plate and axial compression should not be applied.

Lag screw as primary fixation device outside plate

Insertion of lag screw

Insert a lag screw of appropriate length as perpendicular to the fracture plane as possible.

Maintain the reduction by leaving the pointed reduction forceps in situ.

Insertion of lag screw

Neutralization plate

To protect the primary lag screw fixation, apply a precisely contoured plate adjacent to the screw head without axial compression and all screws in the neutral position.

The pointed reduction forceps can then be removed.

Neutralization plate

6. Fixation – lag screw as primary fixation device through plate

Principle

In some fracture configurations, it is advantageous to initially compress the fracture by inserting a lag screw through the relevant plate hole.

If this technique has been used, then axial compression must not be applied secondarily through the plate.

Lag screw as primary fixation device through plate

Insertion of lag screw through plate

Apply a precisely contoured plate to the surface of the anatomically reduced fracture and fix it to one main fragment, either with a neutral screw, or with a bone forceps as illustrated.

Then insert an interfragmentary lag screw through the appropriate plate hole as perpendicular to the fracture plane as possible.

Insertion of lag screw through plate

Insertion of plate screws

Complete the fixation by inserting the plate screws in neutral mode, usually alternating between both main fragments.

Insertion of plate screws

7. Fixation – compression plating with additional lag screw

Insertion of 1st screw

The aim of this maneuver is to create an “axilla” into which the other fragment can be compressed and locked either with a compression screw or using a push-pull technique.

Fix the plate to the appropriate fragment with a screw in neutral mode such that an axilla is created.

Place a reduction forceps on the opposite fragment to hold it in the reduced position against the plate.

Insertion of 1st screw

Insertion of 2nd screw eccentrically

Insert a second screw eccentrically into the reduced fragment.

By tightening the eccentrically inserted screw, axial compression is achieved.

Insertion of 2nd screw eccentrically

Insertion of lag screw

The stability of plate fixation in oblique fractures can be increased by inserting a lag screw through the plate after primary axial compression has been achieved.

Insertion of lag screw

Completed osteosynthesis

Before inserting additional screws check reduction of the fracture and rotation of the forearm.

Additional screws in each fragment are inserted in a neutral position and do not produce additional compression.

Completed osteosynthesis

8. Fixation of wedge fragments

Small wedge fragments

Small wedge fragments that do not have a significant effect on stability should not be addressed. They will become incorporated into the fracture by indirect bone healing.

The two main fragments will be fixed with plating.

Small wedge fragments

Large wedge fragments

Larger wedge fragments, that contribute to the stability of the fixation, may be fixed to one main fragment.

Fixation of the wedge to one main fragment may assist reduction of the residual fragment.

The two main fragments will then be fixed with compression plating either with the lag screw through the plate or separate to the plate.

Large wedge fragments
Go to diagnosis