1. Home
  2. Skeleton
  3. Diagnosis
  4. Indications
  5. Treatment

Authors of section

Authors

Jonas Andermahr, Michael McKee, Diane Nam

Executive Editor

Joseph Schatzker

Open all credits

ORIF - Plate fixation

1. Introduction

The critical step of this procedure is to achieve anatomical reduction as the angle of the glenoid has a direct influence on the functional outcome.

orif plate fixation

To achieve sufficient purchase of the screws in the scapular body, the plate is placed along the lateral border of the scapula.

orif plate fixation

Hardware selection

The standard plate for this fixation is the 2.7 or a 3.5 reconstruction plate.

orif plate fixation

2. Patient preparation

This procedure may be performed with the patient in either a prone position or lateral decubitus position.

3. Approach

Plating of the glenoid neck and scapula is performed through a posterior approach.

approach to the le fort i level of the midface in cleft lip and palate patients

4. Reduction and fixation

Reduction

Reduction may be facilitated by the insertion of one or two K-wires to be used as joysticks for reduction.

orif plate fixation

K-wires are inserted for temporary fixation. Care should be taken that the position of the K-wires does not interfere with the planned position of plate and screws.

The joint block is reduced to restore the anatomical angle of the glenoid fossa. Large deviations of the angle (>20 °) will reduce the functional outcome.

If the reduction is not successful, an alternative procedure where the plate is used as a reduction aid can be employed (see below).

orif plate fixation

A plate of sufficient length should be chosen to allow for the placement of 3 screws in the lateral border of the scapula.

The reconstruction plate is contoured to a near perfect anatomical fit.

orif plate fixation

At least two screws are necessary for sufficient fixation of the glenoid joint block.

Ideally, three bicortical screws are inserted in the lateral scapular border. If the bone stock is poor, locking screws are indicated.

orif plate fixation

5. Alternative reduction and fixation using an angle stable plate

It is best to use a reconstruction plate designed to take angularly stable screws.

The plate is bent at the right spot to subtend the usual angle between the glenoid and the lateral border of the scapula (130 - 135 degrees).

Up to 10 degrees deviation from the anatomically correct angle can be tolerated.

When angularly stable fixation is used the plate does not have to fit the bone, but must fit the overall contour.

orif plate fixation

The plate is fixed to the glenoid segment with two bicortical screws.

If locking screws are used, care must be taken regarding their direction so that they do not enter the joint.

orif plate fixation

The plate is then used as a reduction aid.

Large deviations of the angle (>20 °) will reduce the functional outcome.

A trans-glenoid X-ray is recorded to verify whether the anatomical angle is achieved.

When the correct reduction and the correct angle between the glenoid fossa and the lateral border of the scapula have been achieved, screws are inserted along the lateral border of the scapula.

If indicated, use locking screws and in the articular block beware that they do not enter the joint.

glenoid fossa extraarticular simple

Correct reduction and fixation is verified by image intensification.

orif lag screw fixation

6. Aftercare

The aftercare can be divided into 4 phases:

  1. Inflammatory phase (week 1–3)
  2. Early repair phase (week 4–6)
  3. Late repair and early tissue remodeling phase (week 7–12)
  4. Remodeling and reintegration phase (week 13 onwards)

Full details on each phase can be found here.