Theoretically, the mandibular external fixator could be used to achieve definitive bone healing. Nevertheless, the external fixator does not offer the same degree of stability compared to internal devices.
The external fixator device will often only be applied temporarily, with subsequent replacement by an internal fixation system.
The external pin fixation device gives a high degree of freedom for the frame assembly as the pins can be placed selectively into each segment and connected with short bars to constitute a subunit. Subsequently, the subunits are joined with additional connecting elements to make up the complete framework. In this process, each subunit can be manipulated into a reduced position until the whole construct's final tightening.
Some disadvantages of an external fixator are:
An alternative to the modular technique is the biphasic pin fixation (also known as Joe Hall Morris fixation).
Fracture alignment is first achieved with adjustable connecting rods between the pin pairs (not shown in the illustration).
The aligned pins are then covered with a silicon tube (eg, endotracheal tube, chest tube) and then injected with methyl methacrylate resin.
Alternatively, the pins can be connected with a molded methyl methacrylate block that sets after application.
Finally, after adequate bone healing, the pin fixation and adjustable rods are removed.
This image shows an example of this technique following a self-inflicted gunshot wound, where a plating system or modern fixator was not available.
In this case, the fixation is nonrigid and can lead to instability of the framework. However, it is readily available and may be considered when nothing else is available. Larger threaded Steinmann-type pins would be preferable over K-wires.
The following recommendations apply to optimize the framework stability:
This procedure is typically performed with the patient placed in a supine position.
If the major fragments can be aligned using any of the MMF techniques, MMF is used to simplify the overall assembly process.
The pins are inserted through the soft-tissue envelope overlaying the safe zones into the mandible's lower border.
For more information on the anatomy of the mandible, click here.
Two pins are inserted into each major fragment at an appropriate distance to each other and the adjacent fracture lines.
The length of the threaded portion of the pins is chosen to attain bicortical engagement.
Make a small stab incision to prepare for pin insertion at the predetermined screw locations in the posterior mandible.
The stab incision is done with the blade parallel to the RSTL (relaxed skin tension lines).
Bluntly dissect a soft-tissue channel onto the underlying bone and insert a trocar for soft-tissue protection through the channel until it contacts the bone.
A self-drilling pin is loaded into a handle. Use the trocar as a guide and drive the pin into the bone down to the pin's base.
If a self-drilling pin is not available or not advisable due to fragment instability, make sure to pre-drill before pin insertion.
The clinical photograph shows pin insertion with a trochar.
The two pins in each fragment are connected with a rod and two clamps (as illustrated). The example here shows four subunits.
Apply a connecting rod loosely between two subunits using rod-to-rod clamps.
One fracture is manually reduced by manipulating two subunits, tighten the rod-to-rod clamps
The anterior subunits are linked in the same way as the posterior subunits.
Only one fracture gap is left between two large assembled mandibular portions to connect the anterior and posterior subunits. This gap is reduced and fixed through a connecting rod.
The illustration shows the final assembly of the external fixator using the modular technique.
MMF is removed after the final external fixator assembly to allow for mandibular function.
When a large circumference of the mandible requires external fixation, a pre-contoured rod can be directly attached to the pins.
A template rod is first inserted into the combination clamps.
This template rod is then used to make further adjustments in the bending of the bow-shaped rod.
This clinical picture shows an alternative placement on the inferior border with a history of comminution and multiple infections, and a parasymphyseal defect, after multiple failed attempts at treatment using ORIF and traditional plating techniques.
The postoperative orthopantomogram shows Schanz screws. Note that the screw on the far left is too close to the inferior alveolar nerve.
External fixation of the mandible is usually temporary. Conversion to, and the timing of conversion to internal plate fixation is at the discretion of the surgeon. The general patient condition and local soft tissues must have become suitable to allow this conversion.
The patient should be taught appropriate pin care. Soft tissues around external fixation pins are treated with pin-site cleansing, antibacterial ointments and dressings.
Treatment of intraoral wounds
In the presence of intraoral wounds, the appropriate guidelines for mouth rinses and diet must be followed. Regular appointments at short intervals are mandatory.