An external fixator can be considered for periprosthetic fractures. However, there is an increased risk of introducing an infection to knee arthroplasty. Keeping pins as far away from the arthroplasty as possible will decrease this risk. Consideration of spanning knee external fixation instead of isolated tibial external fixation may also enable the surgeon to protect the knee arthroplasty more effectively.
Throughout this treatment option, illustrations of generic fracture patterns are shown, as four different types:
A) Unreduced fracture
B) Reduced fracture
C) Fracture reduced and fixed provisionally
D) Fracture fixed definitively
The modular external fixator is optimal for temporary use. It is rapidly applied without the need for intraoperative x-rays and can be adjusted later.
Details of external fixation are described in the basic technique for the application of a modular external fixator.
Specific considerations for the tibial shaft are given below.
This procedure is typically performed with the patient in a supine position.
For safe pin placement, use the safe zones and be familiar with the anatomy of the lower leg and the femur.
Any pin placed near a joint should be a minimum of 14 mm away from the joint line to prevent joint sepsis.
In the femur, blunt dissection of the soft tissues and the use of small Langenbeck retractors will minimize muscular damage.
Using a straight clamp, prepare a channel for the insertion of the pin.
Drilling a hole in the thick tibial crest may be associated with excessive heat generation, and there is a risk the drill bit may slip medially or laterally, damaging the soft tissues. As the anteromedial tibial wall provides adequate thickness for the placement of pins, this trajectory is preferable. A trajectory angle (relative to the sagittal plane) of 20-60° for the proximal fragment and 30-90° for the distal fragment is recommended.
Alternatively, to avoid the frame catching on the opposite leg, the pins may be placed more anteriorly. The drill bit is started with the tip just medial to the anterior crest and with the drill bit perpendicular to the anteromedial surface (A). As the drill bit starts to penetrate the surface, the drill is gradually moved more anteriorly until the drill bit is in the desired plane (B). This should prevent the tip from sliding down the medial or lateral surface.
Restore length with a bolster behind the knee to give slight flexion. Please refer to the dedicated section of the proximal tibia in the AO Surgery Reference for additional details.
Thrombo-prophylaxis should be given according to local treatment guidelines.
The pin-insertion technique is important to prevent postoperative complications. Pin care protocol includes:
Local aftercare protocols to prevent pin tract infection are generally in place. The following points are recommended:
In the case of pin loosening or pin tract infection, the following steps need to be taken:
Before changing to a definitive internal fixation, an infected pin tract needs to heal. Otherwise, infection will result.
See patient 7-10 days after surgery for a wound check. X-rays are taken to check the reduction.
The external fixator is usually removed before the definitive internal fixation. Occasionally, it may be maintained to aid in reducing the fracture during plate fixation, but in these cases, there must be no sign of any infection before going to the operating room.