Lag screw fixation

1. Principles

Fracture type

Unicondylar fractures of the proximal phalanx can be transverse, short or long oblique, or comminuted. Typically they are the results of a sports injuries, caused by axial load combined with lateral angulation of the finger.
Condylar fractures tend to be very unstable and should usually be treated operatively. If conservative treatment is attempted, secondary displacement, leading to angulation of the finger, often ocurrs.

Short and long oblique fractures
Short oblique fractures typically originate in the intercondylar notch.
Long oblique fractures more often originate through one of the condyles, splitting proximally towards the diaphyseal cortex on the side of the uninjured condyle.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Caveat

These fractures are rare, but difficult to treat. There is an increased risk of joint stiffness resulting from these fractures.

It is wise to use magnifying loupes in these procedures. Gentle and precise handling throughout the procedure is mandatory.

2. Approach

For this procedure a midaxial approach to the proximal phalanx is normally used.

lag screw fixation

3. Reduction

Anatomical reduction mandatory

Articular fractures must be reduced anatomically. Otherwise, the articular cartilage may be damaged, leading to painful degenerative joint disease and digital deformity.
This illustration shows how even slight unicondylar depression may lead to angulation of the finger.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Visualization of the fracture

In order to gain better a better view of the fracture, use a syringe to irrigate out blood clot with a jet of Ringer lactate.
Gently explore the fracture site to assess its geometry, using a dental pick. The pick can also be used carefully to reduce small fragments. Take great care to avoid comminution of any fragment.
It is important to maintain the vascularity of tiny fragments attached to the collateral ligament, in order to avoid osteonecrosis.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Indirect reduction

Reduction starts with traction in order to restore length.
Lateral pressure, exerted by the surgeon’s thumb and index finger, will then reduce the fracture.
Confirm reduction using image intensification.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Direct reduction of large fragments

A small pointed reduction forceps can be used for larger fragments gently to rock the fracture from side to side. Be careful not to apply excessive force as this can lead to fragmentation.
Confirm reduction using image intensification.

Note
Anatomical reduction is important to prevent chronic instability or posttraumatic degenerative joint disease.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

4. Decision making

Different types of treatment are available for the various fracture types.

Short oblique fractures
Fixation with one lag screw, or a percutaneous K-wire, is recommended.

Long oblique fractures
Fixation with 2 or 3 lag screws is recommended. One or more K-wires is another option, although concerns about instability exist.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Planning screw placement

Large fragments
In large fragments, all screws can be placed safely proximal to the collateral ligament.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Pitfall: Beware of fissure lines

Make sure that the screws are not placed through incomplete fissure fractures.
This is one of the reasons why magnification is important.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Small fragments

If only one screw can be inserted into a small fragment, it will have to be placed within the joint cavity, but through the nonarticular face of the condyle, distal to the collateral ligament.
The lateral aspect of the phalangeal head, which is safe for screw placement, can be approached by flexing the PIP joint.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

5. Small fragment fixation

Screw size

Screw length needs to be adequate for the screw just to penetrate the opposite cortex.
Keep in mind that at the apex of the fragment, the minimal distance between the screw head and the fracture line must be at least equal to the diameter of the screw head. If necessary, a screw of smaller diameter will have to be chosen.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Prepare for drilling

There are two ways to approach the outer surface of the phalangeal head: either by flexing the PIP joint, or with an extended joint, by making a short incision in between the collateral ligament and the accessory collateral ligament.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Drilling

Drill a gliding hole for a 1.0 mm screw, as perpendicularly to the fracture plane as possible, using a 1.0 mm drill bit.
Use a 0.8 mm drill bit to drill a thread hole in the opposite fragment, just through the far (trans) cortex.

Note
Be careful to select appropriately-sized instruments. The use of too large drill bits or screws may result in fragmentation.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Screw length pitfalls

Ensure that a screw of the correct length is used.

  • Too short screws do not have enough threads to engage the cortex properly. This problem increases when self-tapping screws are used due to the geometry of their tip.
  • Too long screws endanger the soft tissues, especially tendons and neurovascular structures. With self-tapping screws, the cutting flutes are especially dangerous, and great care has to be taken that the flutes do not protrude beyond the cortical surface.
Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Insert lag screw

Insert the lag screw and gently tighten it to compress the fracture.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Pitfall: protruding screw head

A protruding screw head can cause ligament irritation and eventual joint stiffness.
In order to avoid such protrusion, slightly enlarge the entrance to the screw hole in order partially to sink the screw head. Do not attempt to countersink with too large a tool, and be careful not to overtighten the screw, to avoid fragmentation.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

6. Large fragment fixation

Preliminary fixation (large fragments)

Long oblique fractures can be preliminarily fixed by inserting a K-wire. Be careful to place it in such a way that it will not conflict with later screw placement.

Avoid inserting a K-wire into small fragments, as they are in danger of fragmentation.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Drilling and alternative preliminary fixation

Leaving the reduction forceps in place, drill a gliding hole as perpendicularly to the fracture plane as possible, using a 1.5 (or 1.3) mm drill bit for a 1.5 (or 1.3) mm screw. Insert a 1.5 (or 1.3) mm drill sleeve into the gliding hole.
Use a 1.3 (or 1.0) mm drill bit to drill a thread hole in the opposite fragment, just through the far (trans) cortex.
Leave the drill bit in the drill hole, preliminarily to hold the reduction if no K-wire has already been used for this purpose.
Remove the reduction forceps.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Screw size selection

The exact size of the diameter of the screw used will be determined by the fragment size and the fracture configuration.
The various gliding hole and thread hole drill sizes for different screws are illustrated here.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Pitfall: Oblique measuring

When measuring for screw length in oblique drill holes, the measurement to the acute angle is different from the measurement to the obtuse angle. This problem increases with the degree of obliquity.
Always measure both angles and use the longer measurement. However, keep in mind that too long a screw can protrude to the extent that it puts the soft tissues at risk.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Screw length pitfalls

Ensure that a screw of the correct length is used.

  • Too short screws do not have enough threads to engage the cortex properly. This problem increases when self-tapping screws are used due to the geometry of their tip.
  • Too long screws endanger the soft tissues, especially tendons and neurovascular structures. With self-tapping screws, the cutting flutes are especially dangerous, and great care has to be taken that the flutes do not protrude beyond the cortical surface.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Lag screw insertion

Countersinking in diaphyseal bone
There are two important reasons for countersinking:

  1. The risk of soft-tissue irritation is greatly reduced by ensuring that the the screw head protrudes only minimally from the bone surface.
  2. Countersinking ensures that the screw head has a maximal contact area with the bone, distributing the forces from the screw head more widely and more biodynamically than an unsunk head.
Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Pitfall: breaking the cortex in diaphyseal bone

Do not advance the countersink too deeply into the cortex: the cortical thickness will determine the depth of countersinking. Excess penetration risks break-through of the screw head when tightened and loss of fixation. Countersinking is, therefore, done by hand and not with a power tool.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

No countersinking in the metaphysis

Do not countersink the screws in the metaphysis as its cortex is very thin.
If countersinking is attempted, all purchase and compression may be lost due to screw-head breakthrough.

Do not countersink the screws in the metaphysis as its cortex is very thin.

Preliminary fixation with drill bit

The drill bit can be left in the drill hole for preliminary fixation. This saves space, as no K-wire is needed.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Drill for proximal screw

Towards the proximal apex of the fracture line, drill a gliding hole for a second lag screw. This screw, too, should be placed as perpendicularly to the fracture plane as possible, using a 1.5 (or 1.3) mm drill bit for a 1.5 (or 1.3) mm screw. Insert a 1.5 (or 1.3) mm drill sleeve into the gliding hole.
Use a 1.0 (or 0.8) mm drill bit to drill a thread hole in the opposite fragment, just through the far (trans) cortex.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Insert distal screw

Insert the distal lag screw. Do not completely tighten it at this time. The screw should penetrate the opposite cortex.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Insert proximal lag screw

Now insert the proximal lag screw. This screw, too, should penetrate the opposite cortex.
Alternate tightening of the two lag screws helps to avoid tilting of the fragment and applies even compression forces over the whole fracture surface.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Check using image intensification. The reduction must be anatomical.

Check stability of the fixation by passive flexion and extension of the PIP joint, and by applying gentle lateral and rotational motion. This will help to determine stability in order to establish strategies for rehabilitation.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

Caveat: Changing fracture plane

Hastings and Weiss described a fracture type in which the fracture plane changes between the condylar and metaphyseal zones.
In such a fracture configuration, it is important to observe that all the lag screws are inserted as perpendicularly to the local fracture plane as possible.
Be careful to confirm correct fracture planes under direct vision and in different radiographic views.

Oblique distal condylar fracture of the proximal phalanx – Lag screw fixation

7. Aftertreatment

Postoperatively

Protect the digit with buddy strapping to the adjacent finger, to neutralize lateral forces on the finger.

/Surgery/70/70_7_Reha_2
While the patient is in bed, use pillows to keep the hand elevated above the level of the heart to reduce swelling.
k wire fixation

Functional exercises

The patient can begin active motion (flexion and extension) immediately after surgery.

/Surgery/70/70_7_Reha_1

For ambulant patients, put the arm in a sling and elevate to heart level.

Instruct the patient to lift the hand regularly overhead, in order to mobilize the shoulder and elbow joints.

/Surgery/70/70_16_reha_2

Follow-up

See the patient 5 days and 10 days after surgery.

Implant removal

The implants may need to be removed in cases of soft-tissue irritation.

In case of joint stiffness, or tendon adhesion’s restricting finger movement, tenolysis, or arthrolysis become necessary. In these circumstances, take the opportunity to remove the implants.